RA RA 混凝土共振分析仪)
Non-destructive, QuickDetermination of Compressive Strength and Modulus of Elasticity onSamples of Cement, Mortar and Concrete
Widerange of strength – Mortar, normal concrete up to Ultra HighPerformance Concrete RA Concrete
Higher level of production monitoring and betterquality assurance Easier and more precise than ultrasoundmeasurements
Measurement only takes a fewseconds
Robust algorithms
Noalignment effort
Mobile use – up to 6 hours batteryoperation
Validity of measurement immediatelyidentifiable
ongoing recording of different factorsinfluencing the strength development: Temperature, moisture,composition, additives, aftertreatment, freeze-thaw cycles,chemical and other environmental impacts starting from the encasingdate of the young concrete.
ForLaboratory, Production and Construction Site Measurements onCylinder, Bar and Cube Samples
Bymechanical stimulation of resonance frequencies, the RA100 Concretemeasures according to the Impact Resonance Method easily andquickly the sound velocity. Compressive strength, static anddynamic modulus of elasticity are determinated directly and withcalibration values.
Test Bench
•High quality steel - stainless steel design
•Dimensions: 530 x 250 x 230 mm
•Weight: 7 kg • Max. load up to 100 kg
•Mount for adapter and chuck • Acrylic glass downpipe with steelballs
•Integrated microphone 50 Hz – 18000 Hz
• Test bench operating temperature -10 to+85 C°
• Sample temperature range -40 to +150 C°(temporary up to +250 C°)
Computer andSoftware
•Standard 10 Inch Tablet (opt. rugged Toughpad)
•Windows 8.1
• External keyboard
•USB connector cable
•Pre-installed RA100 Concrete software
•Extensive possibilities for documentation andarchiving
•Interface for easy data export (CSV, DIFF etc.)
Measurement Range
•Compressive strength: 0,1 MPa - 200 MPa
•Modulus of elasticity: 1 - 70 GPa
• Sound velocity: 300 m/s - 6000m/s
Referenceto Standards
•Resonance measurements of concrete test specimen for determinationof dynamic modulus of elasticity are described in American ASTMC215, British BS 1881-209
•Dynamic modulus of elasticity can be converted to static modulus ofelasticity by established reference curves. Direct measuring ofstatic modulus is standardized in European EN 12390-13 and AmericanASTM C469
•Compressive strength can indirectly be determined by calibrationwith the measured sound velocity. Standardized measurement ofcompressive strength are usually made according to European EN12390-3 (cube, cylinder), EN 196-1 (mortar prism) and American ASTMC39 (cylinder)
• Inthe freeze-thaw resistance test standards European CEN/TR 15177 andAmerican ASTM C666 the relative change of the dynamic modulus ofelasticity is used to evaluate internal structuraldamage.
Samples
•Test bench for common standard test samples of the concreteindustry: Cylinders 300 x 150 mm, cubes 150 mm and bars 40 x 40 x160 mm
•Moreover, all bar-shaped samples like drill cores with uniformprofiles (cylinders and bars) with 1 - 16 cm diameter and 10 - 40cm height
Ein Probekörper wird durch einen definierten, mit einer Stahlkugeleingeleiteten Impuls zur Schwingung angeregt. Währenddessen erfolgteine Messung des Frequenzgangs im Hörschallbereich und daraus eineBerechnung der Resonanzfrequenz. Auf Grundlage der Resonanzfrequenzund der Probekörperabmessungen, wird die longitudinaleSchallgeschwindigkeit clong berechnet. Der dynamischeElastizitätsmodul Edyn ergibt sich anschließend als Produkt aus demQuadrat der Schallgeschwindigkeit und der – als Eingangsgröße zuimplementierenden – Rohdichte. Zu beachten ist, dass dieserZusammenhang nur dann zu präzisen Ergebnissen führt, wenn dasVerhältnis der Schallwellenlänge zum Durchmesser des Probekörpersrelativ groß ist. Dieser Verhalt ist bspw. bei metallischenRundstäben, Mörtelprismen und sogar bei Betonwürfeln und -zylindernmit einer Kantenlänge von bis zu 15 cm gegeben. Die Höhe derProbekörper sollte zum Zweck einer präzisen Bestimmung derSchallgeschwindigkeit mindestens 10 cm betragen. Geprüft werdenkönnen Stoffe unterschiedlichster Dichte, von Dämmstoffen übermineralische Stoffe bis hin zu metallen. Über clong und Edyn hinausist eine Bestimmung des Dämpfungsgrads D möglich.